Lalwani Balasingam Olazarra Saffari

From Psych 221 Image Systems Engineering
Jump to navigation Jump to search

Introduction

This quarter, our team worked on characterizing fluorophores purchased by the Valdez Lab in the Department of Otolaryngology, Head and Neck Surgery at the Stanford School of Medicine to determine optimal materials for cholesteatoma visualization. The fluorophores we measured include elastin, keratin, NADH, and collagen. Our inspiration to pursue a project focused on using fluorescence to better image cholesteatomas was inspired by work being done in the Valdez lab as well as oral imaging work being done by Dr. Joyce E. Farrell and Zheng Lyu in Stanford's Department of Electrical Engineering. We hope our work can shed some light on which fluorophores may be best for producing a label-free method in congenital cholesteatoma detection via autofluorescence microscopy. To aid these efforts, our team worked on characterizing and analyzing such fluorophores to determine the optimal selections for clinical use given their respective absorption and emission spectra. Following these measurements, we designed a fluorophore matrix that accommodated our selected fluorophores, which enabled further measurements and characterization to be performed using the OralEye imaging system developed by Dr. Farrell and Zheng. While our group focused on using this matrix for measurements of our chosen fluorophores, the matrix was designed to be customizable for users by allowing them to insert and remove their own fluorophores of interest. While our project allowed us to gain a great deal of information about our selected fluorophores and expand upon our own understanding of how light interacts with materials, we plan on refining our methods and taking further measurements in the year to come.

Background

In this section, we will provide both the clinical and technical background that motivated us to pursue this project. Additionally, we will discuss the reasons we deemed fluorescence to be an ideal diagnostic tool given our clinical objective.

Clinical Background

Cancers of the head and neck can be life-altering and even deadly, as well as difficult to detect early. While the 5-year survival rate is high, these cancers can often go unnoticed until a very late stage [12]. As a result, there is a strong and urgent need for better tools to detect head and neck cancers.

We were driven to find a label-free method for autofluorescence imaging of cholesteatomas. Cholesteatomas are non-malignant, non-cancerous growths that typically form behind the tympanic membrane [1, 2]. They are similar to cysts, and often occur in people who are prone to otitis media (or middle ear infections), and many are in fact congenital [2, 3]. Our objective is to create a label-free method to characterize biomolecular changes in the keratinized epithelial tissue by exploiting the molecular specificity of autofluorescence spectroscopy. Specifically, we will measure the spectral signatures of biological molecules like NADH and elastin, both of which are naturally occurring in the body and can fluoresce upon absorption of ultraviolet (UV) or visible light. Ultimately, being able to quickly and reliably visualize these cholesteatomas is desirable for early diagnosis and complete resection of the lesions, in order to better preserve the patient’s hearing ability.

Technical Background

Fluorescence occurs when a molecule absorbs a high energy (short wavelength) light and promotes an electron to a higher energy level; subsequently, the electron will fall back to its ground state and re-emit a photon with an energy equal to the difference between the two energy levels. The energy of this photon is lower than that of the excitation light, meaning that the emitted light will have a longer wavelength. A large number of biomolecules, such as NADH, absorb light in the UV range and emit light in the visible spectrum. The phenomenon of absorbing a shorter wavelength light followed by re-emitting a longer wavelength light is called fluorescence, and each molecule has its own characteristic fluorescence signature.

To better visualize neoplastic regions, researchers have focused on creating optical imaging methods that exploit various interactions between light and tissue [4]. For our purposes, we are interested in the autofluorescence of tissue behind the tympanic membrane. Autofluorescence is a product of the numerous fluorophores in the middle ear cavity (such as collagen and elastin), and is particularly sensitive to changes in tissue morphology and biochemical alterations of neoplasia [5, 6]. Therefore, tissue autofluorescence is being harnessed as a diagnostic tool for oral cancers, various head and neck lesions, and for our purposes, in cholesteatomas. By exciting fluorophores in the ear with UV or even visible light, the emitted light will fluoresce at longer wavelengths. We can measure this emission spectra following excitation and, with the use of an absorbing filter to block reflected illumination, can visualize the fluorescence with our eyes [4]. For our purposes, we will not be visualizing these fluorophores with our eyes so much as we will be using a fluorometer to accurately measure the emission spectra and record the peak wavelength following excitation.

Methods

We acquired four fluorophores (NADH, collagen, elastin, and keratin), then measured the fluorescence signal of each using a fluorometer and the OralEye camera, which was provided by Dr. Farrell. The purpose of our initial fluorescence measurements, which were performed using a fluorometer, was to determine whether we could measure emission spectra that were consistent with those recorded in the literature for each of our fluorophores. Secondary measurements, which were performed using the OralEye camera, sought to determine whether we were able to effectively measure similar emission spectra using the 385 nm excitation light source associated with the OralEye camera.

Fluorometer Measurements with Haribo's Fluorolog Device

The fluorometer we used for the measurement of the fluorophores was made available to us by the Stanford Department of Chemistry (machine located in Stauffer 1). Figure 1 is an image of the Fluorolog machine we used to measure the emission and absorption spectra of the fluorophores.



Figure 1: Fluorometer, Fluorolog by Haribo.


The fluorometer’s excitation wavelength ranges from 250 nm to 400 nm, while the emission spectrum has a range of 250 nm to 1000 nm. Therefore, this tool would be particularly useful in our measurements, as its dynamic range met the specifications required by our fluorophores.

As for preparing the fluorophores themselves, we kept the concentrations consistent within each cuvette in order to effectively compare spectra between NADH, elastin, collagen, and keratin. For our liquid samples (keratin and elastin, which was mixed with deionized (DI) water), we put ~30 ul into 1.5 mL of DI water in a clear cuvette. For NADH, which came in the form of small "dip-n-dot"-like samples, we placed between 3-5 dots (depending on the size of the dots) in 1.5 mL of DI water in the cuvette. Finally, for collagen, which arrived as a cotton-ball-like material, we evenly divided the sample into 5 pieces and placed 1 piece in the cuvette (with 1.5 mL of DI water) for spectral measurements. Based on this concentration level, we set the Fluorolog’s slit size to 3.0 mm, in order to produce emission spectra that did not have disproportionately high intensities. As a reference, concentrated solutions are typically those that look cloudy in the cuvette; and those solutions that are more heavily concentrated must have a smaller slit width in order to appropriately measure the fluorescence at a reasonable intensity.

Measurements with the OralEye Camera

To determine if we could measure fluorescence using the OralEye camera, we built a matrix that not only accommodated our water-soluble fluorophores, but also accommodated standard fluorescent microscope slides (and similarly-prepared, alternative fluorophores identified by the user). A three-dimensional rendering of the matrix, which was modeled after standard color calibration targets, is pictured in Figure 2. Performing measurements using the fluorescent calibration targets allowed us to determine whether we were able to measure any fluorescence at all, while performing similar measurements using our chosen fluorophores allowed us to attempt to measure and compare their respective fluorescence signals with the results obtained using the fluorometer.


Figure 2: A three-dimensional rendering of the matrix we designed for our OralEye fluorescence measurements.


Prior to performing the fluorophore fluorescence measurements, we prepared four standard glass microscope slides: one for each of the respective fluorophore solutions. Sample preparation varied depending on the state in which the fluorophores were provided, but was largely consistent with the sample preparation described in the fluorometer measurements section. For keratin and elastin, which were provided in liquid form, we dissolved approximately 20 uL of the fluorophore in 1 mL of DI water. For NADH and collagen, the concentrations were slightly less straightforward. The NADH provided took the form of a collection of non-uniform beads; to prepare this sample, we dissolved approximately five of the beads in 20uL of distilled water. Finally, the collagen was presented as a cotton-ball-like substance; to prepare this sample, we tore off a small portion of the collagen substance, then dissolved it in 20uL of distilled water. After the fluorophore solutions were prepared, we sealed them by placing glass cover slides on top.

Figure 3: A schematic describing the set-up used for our OralEye fluorescence measurements.


Our four fluorophore samples were then installed in the aforementioned matrix in order to perform fluorescence measurements using the OralEye camera. We fixed the matrix to an optical stand, which was then fixed to an optical table. The optical stand allowed us to move the matrix vertically in order to capture data corresponding to both the top and bottom row samples. The OralEye camera was installed on a track that allowed us to move the camera/light source horizontally in order to illuminate and capture data corresponding to both the left and right column samples. In order to actually capture the emission spectra, we situated a spectroradiometer approximately one meter from the matrix. Because the spectroradiometer, alone, does not have a dynamic range high enough to distinguish a weak emitted fluorescence signal from a strong reflected light signal, we equipped the spectroradiometer with a Y44 longpass filter. By installing this filter, we were able to filter out a large fraction of the signal produced due to light reflected from the 385 nm light source. (Further discussion regarding the efficacy of this filter with respect to our particular results can be found in the “Results” and “Conclusions” sections.)

Results

Results from Fluorometer Measurements

We found the following spectra for the NADH, collagen, elastin, and keratin fluorophores using Haribo’s Fluorolog. Additionally, we chose to excite these fluorophores at specific wavelengths based on work being conducted in the Valdez lab and methods used in previous literature [4, 7, 8, 9].

NADH

Figure 4 summarizes the absorption and emission spectra we recorded for NADH at various specified wavelengths.

Figure 4: The leftmost plot illustrates the absorption spectrum of NADH at an emission of 415 nm. The following two plots (in the middle and the right) are emission spectra where NADH is excited at 293 nm and 317 nm, respectively. Note that, on the rightmost emission spectrum, the concentration of NADH was doubled before being excited at 317 nm.

These results are consistent with those reported in previous literature [7]. Out of all the spectra recorded from fluorophores, NADH was the most consistent with previously recorded results.


Collagen

Figure 5 summarizes the absorption and emission spectra we recorded for collagen at various specified wavelengths.

Figure 5: The leftmost plot illustrates the absorption spectrum of collagen at an emission of 415 nm. The following two plots (in the middle and the right) are emission spectra where collagen is excited at 317 nm and 293 nm, respectively. On the rightmost emission spectrum, the collagen concentration was halved before being excited at 293 nm.

These results are not consistent with results reported in previous literature, with our spectrum appearing left-shifted when compared with reported literature spectra [8]. The reason behind these readings being at a lower wavelength is discussed in greater detail in the troubleshooting section below.


Elastin

Figure 6 summarizes the absorption and emission spectra we recorded for elastin at various specified wavelengths.

Figure 6: The leftmost plot illustrates the absorption spectrum of elastin at an emission of 415 nm. The following two plots (in the middle and the right) are emission spectra where elastin is excited at 325 nm. Note that, on the middle emission spectrum, the concentration of elastin was halved before being excited at 325 nm.

These results are nearly, though not completely, consistent with results reported in previous literature [8]. While the absorption spectrum is relatively close to what has been seen in previous findings, the emission spectrum for elastin typically is more prominent at around 400 nm. The rightmost emission spectrum is closer to this value, but is just a hair shy. Additionally, these peaks are broader than typical spectra for elastin.


Keratin

Figure 7 summarizes the absorption and emission spectra we recorded for keratin at various specified wavelengths.

Figure 7: The leftmost plot illustrates the absorption spectrum of keratin at an emission of 415 nm. The following two plots (in the middle and the right) are emission spectra where keratin is excited at 293 nm and 415 nm, respectively. Note that, on the rightmost emission spectrum, the concentration of keratin was doubled before being excited at 415 nm.

These results are nearly, but not completely consistent with results from previous literature [9]. While the absorption spectrum is relatively close to what has been shown in previous findings at high keratin concentration, the emission spectrum for keratin is typically more prominent at approximately 390 nm. Additionally, we also see a secondary, smaller peak at around 360 nm. We explain the reasons behind this error in more detail in the troubleshooting section.


Troubleshooting the Measurements and Adjusting our Protocol

One of the main reasons why our spectra may differ from those reported for the same fluorophores in previous literature may be due to cross-contamination of the fluorophores. Before beginning our experiment, we were prepared to measure our samples using cuvettes with beveled edges. These cuvettes produced a great deal of scattering, so we switched to using clear cuvettes without any beveled edges. This technique allowed us to eliminate the presence of Raman peaks for samples measured in the beveled cuvettes; therefore, switching over to the clear cuvettes proved to be the best solution for eliminating scattering. Unfortunately, we did not have enough of these clear cuvettes for each sample, so we had to re-use one clear cuvette for numerous samples. While we tried our best to clean and remove all fluorophore remnants from previous trials, we would be remiss to not acknowledge the potential for any lingering fluorophores to interfere with the measurements of subsequent fluorophores.

Raman peaks can be detected by changing the excitation spectrum by a few nanometers and observing if there is a shift in the emission peak by the same amount. If there is such a shift, then we knew that we were witnessing a Raman peak. This was how we identified the cuvette type as being the main culprit for our scattering problems. Raman peaks generally occur close to the excitation wavelength; and by implementing a low pass filter cut-off above the Raman peak, we can make sure that we are measuring the true signal of the fluorophore.

In addition to this Raman scattering (as seen in the case of the beveled cuvette), we also needed to address the second-harmonic light arising from the excitation source. The second harmonic excitation is when light from the excitation source (a white light with a grating filter on top) measures not only the wavelength of interest (for example, 300 nm) but also the second harmonic of that wavelength (600 nm). The second harmonic is then recorded as a very strong and narrow peak. To avoid this issue, we made sure that the high pass filter in the Fluorolog's fluorescence settings is set to 10 nm below the second harmonic.

Another reason why some of the emission spectra may be shifted from those recorded in the literature may be due to not exciting the fluorophores at ideal wavelengths. In particular, we are most interested in redoing these measurements by exciting the fluorophores at a wavelength closer to the visual spectrum: 385 nm.

While these potential sources of error could have skewed some of our measurements, they are relatively easy to fix by ordering a full panel of clear cuvettes and adjusting our excitation wavelength on the Fluorolog.

Results of OralEye Camera Measurements

In order to understand the data we collected for our fluorophore measurements, we plotted the emission spectra corresponding to each of our fluorophores as a function of wavelength. This plot can be seen in Figure 8. In inspecting this plot, we made two primary observations. First, we noted that all four of our fluorophores demonstrated a sharp increase in measured fluorescence at a wavelength of approximately 425 nm, peaking shortly thereafter. The fact that all four of our fluorophores peaked in this region was not necessarily worrisome by itself, as we expected very little signal below 425 nm due to the presence of the Y44 filter, which should ideally only pass light corresponding to wavelengths greater than 425 nm. As such, it seemed reasonable that we would see a spike in all four of the signals at wavelengths beyond this threshold. The second observation we made was that the peak magnitude of the measured fluorescence for three of our four fluorophores was nearly identical (exact values for maxima are specified in the legend of Figure 8). This feature of the plot was slightly more concerning, as it suggested the possibility that we were measuring the same signal for all of our fluorophores despite the fact that we expected different emission spectra. This raised concerns regarding the integrity of our data, and pointed to the possibility that we measured reflected light rather than a true fluorescence signal.


Figure 8: A plot of the emission spectra we measured for our selected fluorophores using the OralEye camera set-up.


In an effort to better understand our results, we plotted the emission spectra of our fluorophores alongside the filtered spectrum of the 385 nm light source (see Figure 9). That is, we multiplied the spectrum of the 385 nm light with the transmittance of the Y44 longpass filter, then compared the resulting spectrum with the emission spectra of our fluorophores. By plotting these together, we were able to assess how the magnitude of the filtered, reflected light compared with the magnitude of the measured fluorescence signal corresponding to each of our fluorophores. As Figure A.A indicates, even after the transmittance profile of the Y44 longpass filter was applied, some amount of the reflected 385 nm was transmitted. Furthermore, the amplitude of the reflected light signal was visibly larger than the amplitude of any of the fluorescence signals at 425 nm. This suggested to us that there was, in fact, a strong possibility that our fluorescence measurements were significantly distorted by light reflected from our light source.


Figure 9: Our fluorophore emission spectra plotted alongside the filtered, 385 nm reflected light spectrum. (Note: The filtered spectrum was acquired by multiplying the unfiltered 385 nm reflected light spectrum that we measured by the expected transmittance of the Y44 filter.)


In order to further explore the possibility that we had not measured a legitimate fluorescence signal for our fluorophores, we plotted the normalized fluorescence emission spectra for all four of our fluorophores and compared them with previously-established literature values. Figure 10 summarizes the emission spectra results obtained by Yang, et. al. while Figure 11 shows the normalized emission spectra for each of the four fluorophores we measured. While we would not expect the exact magnitude of our fluorescence measurements to match those obtained by Yang, et. al. due to different normalization criteria and excitation spectra, we would expect to see the same trend in terms of relative magnitudes of measured fluorescence at a given wavelength. For example, at a wavelength of 500 nm, we would expect to see (in order of lowest magnitude response to highest magnitude response): collagen, elastin, and NADH, with elastin and NADH showing significantly higher fluorescence signals than collagen. However, our results show (in the same increasing order): elastin and collagen (where elastin and collagen are nearly indistinguishable at 500 nm), then NADH. This discrepancy in our values versus previously established literature values further supported our hypothesis that we measured reflected light rather than fluorescence.


Figure 10: Yang, et. al.’s results for the emission spectra of collagen, elastin, and NADH [8].


Figure 11: A plot summarizing the normalized emission spectra that we measured (using the OralEye camera) for collagen, NADH, keratin, and elastin)


After establishing that we did not measure a robust fluorescence signal for our fluorophores, we wanted to determine whether or not we were able to effectively measure the stronger, more robust fluorescence signal associated with our fluorescent calibration slides. This would help us to clarify whether our undesirable fluorophore measurements were rooted in the weak fluorescence of those fluorophores, or a general inability to measure fluorescence using our previously-described experimental configuration and measurement schema. Just as we had done in our fluorophore measurement analysis, we began by plotting the emission spectra of the calibration slides alongside the filtered reflected light spectrum. From this initial plot, pictured in Figure 12, it immediately became clear that the fluorescence signal associated with the calibration slides was far more robust when compared with the signal associated with the reflected light than the fluorophores were. That is, the magnitude of the fluorescence signal for the calibration slides was much larger than the magnitude of the signal associated with the reflected light at the wavelengths we were interested in. This indicated that the signals we were measuring at or above 425 nm were likely due to actual fluorescence rather than reflected light. As such, we concluded that our experimental configuration could, in fact, measure fluorescence provided that the fluorescence produced by the sample was greater in magnitude than that of the reflected light.

Figure 12: The emission spectra we measured for the calibration slides plotted alongside the filtered, reflected light signal. (Note: All measurements in this plot were performed using the OralEye camera experimental configuration.)

To further assess the integrity of the calibration slide measurements, we compared the normalized emission spectra with the emission spectra provided by the slide manufacturer. Figure 13 shows a plot of the normalized emission spectra of the fluorescent calibration slides after the Y44 filter transmittance profile was applied. When we compare this with the expected results published by the slide manufacturer, which are summarized in Figure 14, we saw fairly consistent results for all of the calibration slide emission spectra except for the results associated with the blue slide. While the manufacturer results indicate a blue peak at approximately 375 nm, our blue peak occurs at approximately 458 nm. While we are not entirely sure about why we saw this inconsistency, one reason might be that the expected blue emission peak occurs below 425 nm, which is the Y44 filtering threshold. As such, we believe that we may have been unintentionally filtering out a portion of the blue calibration slide’s emission spectrum, thus distorting the results.


Figure 13: A plot of the normalized emission spectra associated with our fluorescent calibration slide measurements.



Figure 14: A plot of the expected emission spectra for our fluorescent calibration slides (as provided by ThorLabs) [10].

Conclusions

Throughout our work on this project, particularly during the OralEye measurement analysis, we became acutely aware of the challenges associated with measuring weak fluorescence signals in the presence of even the smallest quantities of reflected light. Despite our efforts to diminish the effects of light reflected by our light source by implementing a Y44 longpass filter, we still saw sufficient enough transmission such that our fluorophore fluorescence signal was indistinguishable. While we regret that we were not able to measure a more robust fluorescence signal for our selected fluorophores, we feel encouraged by the promise we saw in our calibration slide measurements, and feel confident that meaningful fluorescence measurements can be collected if the issues of reflected light leakage are addressed going forward. For recommendations regarding next steps for this particular portion of the project, please see the “Future Work” section below.

Future Work

In the future, we plan to evaluate the effects of different concentrations on our ability to distinguish a fluorescence signal from the background reflected light signal. Our experiments with the Fluorolog fluorometer indicate that concentration can have a noticeable effect on our ability to measure fluorescence, in addition to the excitation wavelength itself. Assuming a similar dependency on concentration in our OralEye camera set-up, we would like to perform many experiments to find the “sweet-spot” concentration value that gives us measurable fluorescence. By the same token, we would like to perform more measurements on the fluorometer by exciting our fluorophores at the 385 nm wavelength, to better compare the measurements between the OralEye Camera and the emission spectra. This will also help us find emission spectra that are consistent with those recorded in previous literature.

In order to acquire more meaningful data using the OralEye camera set-up, we must address the issue of measuring weak fluorescence in the presence of a relatively strong reflected light signal. One obvious way to do this is to implement a more effective filter in order to prevent excessive reflected light from “leaking” into our fluorescence measurements. One possibility would be to apply a filter with an even higher threshold, such as a Y48 or Y50 filter (with cut-off wavelengths of ~480 nm and ~500 nm, respectively), to further reduce the impact of the light reflected from the 325 nm light source. As mentioned earlier, another possibility would be to increase or decrease the concentration of the fluorophore samples we measure; doing so might increase the signal produced by the fluorophores, and thus reduce the issues associated with the weak signal we saw in our earlier measurements. By exploring these sorts of solutions, we feel confident that a robust set of measurements can be acquired for our selected fluorophores.

References

[1] E. L. Derlacki and J. D. Clemis. “Congenital cholesteatoma of the middle ear and mastoid.” Ann. Otol. Rhinol. Laryngol. 1965, 74, 706.

[2] M. J. Levenson, S. C. Parisier, P. Chute, S. Wenig and C. Juarbe. “ A review of twenty congenital cholesteatomas of the middle ear in children.” Otolaryngol. Head. Neck. Surg. 1986, 94, 560.

[3] G. T. Richter and K. H. Lee. “Contemporary assessment and management of congenital cholesteatoma.” Curr. Opin. Otolaryngol. Head. Neck. Surg. 2009, 17, 339.

[4] Shin D, Vigneswaran N, Gillenwater A, Richards-Kortum R. “Advances in fluorescence imaging techniques to detect oral cancer and its precursors.” Future Oncol. 2010, 6, 7: 1143–1154. doi:10.2217/fon.10.79

[5] De Veld DC, Witjes MJ, Sterenborg HJ, Roodenburg JL. “The status of in vivo autofluorescence spectroscopy and imaging for oral oncology.” Oral Oncol. 2005, 41: 117–131.

[6] Roblyer D, Richards-Kortum R, Sokolov K, et al. “Multispectral optical imaging device for in vivo detection of oral neoplasia.” J Biomed Opt. 2008, 13, 2: 024019.

[7] Becker-Hickl. Metabolic Imaging by NAD(P)H and FAD Film. Web: https://www.becker-hickl.com/applications/metabolic-imaging/. Accessed December 2019.

[8] Pu, Yang et al. “Changes of collagen and nicotinamide adenine dinucleotide in human cancerous and normal prostate tissues studied using native fluorescence spectroscopy with selective excitation wavelength.” Journal of Biomedical Optics 2010, 15, 4. Print.

[9] Pena, A.M. et al. “Spectroscopic analysis of keratin endogenous signal for skin multiphoton microscopy.” Abstract, Laboratory for Optics and Biosciences, CNRS/INSERM, Ecole Polytechnique: 2005.

[10] ThorLabs. “Fluorescent Microscope Slides and Alignment Disks.” Web: https://www.thorlabs.com/newgrouppage9.cfm?objectgroup_id=12142. Accessed December 2019.

[11] Joyce Farrell and Brian Wandell 2019, ISETCam, "https://github.com/ISET/isetcam".

[12] Oral Cancer Foundation. "Oral Cancer Facts." Web: https://oralcancerfoundation.org/facts/?fbclid=IwAR3Ko_VzgFl5OPTKAy6jFuTtw0adqB0ZIw8EEprdzif7s9-6Z1nAlWBn_J4. Accessed December 2019.

Appendix I

ISETCam ISETFluorescence was used to analyze measurements from the OralEye Camera [11].

Appendix II

Anand and Persiana worked on characterizing the fluorophores by measuring the absorption and emission spectra of each fluorophore. They used a fluorescence spectrometer (fluorometer) in Stanford's Department of Chemistry that can measure wavelengths from the UV range to 1,000 nanometers. Additionally, Haribo's Fluorolog fluorometer has a sub-nanometer image resolution. Following fluorimeter training, preparation of samples, and recording of measurements, Anand and Persiana analyzed the data and interpreted it for the writing of this report and the presentation.

Based on the results of the fluorophore absorption and emission spectra, Sofie and Ramya constructed a matrix containing these water-soluble fluorophores. The design was modeled after that of a Connect Four, and was 3D-printed at Stanford in the Huang building's basement. Following construction of the matrix, Sofie and Ramya proceeded to take measurements with the OralEye Camera in Dr. Farrell’s lab. They then subsequently plotted their results using MATLAB scripts found in Appendix I.

We all contributed to the final in-class presentation and our Wiki page submission.

Acknowledgements

We would like to express our appreciation for the countless hours that Professor Brian A. Wandell, Dr. Joyce E. Farrell, and Zheng Lyu spent on making this course truly exceptional. Across our team, the class has inspired several new interests, including presenting magneto-sensor research at the next imaging symposium, exploring a career in ophthalmology, and continuing to refine our fluorophore results in the year to come. We are tremendously grateful for your time, energy, and investment in our team. It brought us great joy and honor to be your students.